FAQ

  • What are the main differences between 4100 Series and 4300 Series LCR Meters?
    4300 includes a LAN (Local Area Network) interface as standard 4300 supports up to 9-bin categorization in the Bin Handler mode, capable of controlling multiple binning operations in production environment 4100 supports only 2-bin categorization: Pass and Fail
  • What is purpose of trimming?
    The purpose of trimming is to eliminate the effects of stray capacitance or series impedance in the connecting leads or fixture.
  • At what situations is trimming necessary?
    When the lead set or fixture is changed; when the highest possible accuracy is required for measurements of very high or low impedances; and when the instrument is switched between 4-terminal and 2-terminal operation.
  • How often does the test equipment need the calibration by the manufacturer?
    Wayne Kerr recommend the test equipment be calibrated by the manufacturer every six months.
  • What is the procedure to have my test equipment calibrated by the manufacturer?
    You can either call the service center or email to service@waynekerr.com. one RMA number will be issued to each test equipment. Send the test equipment to the service center with the RMA number.